Faster SafeFAST Classic vertical laminar flow cabinets are Class II Microbiological Safety Cabinets - designed and built to performance requirements of the EN-12469:2000 European Standard and NSF/ANSI 49 American Standard, with 70% of the air re-circulated via the main H14 HEPA/ULPA filter according to EN 1822 within the cabinet, whilst the remaining 30% is discharged through an exhaust H14 HEPA/ULPA filter.
External structure in epoxy powder coated cold-rolled steel for excellent corrosion resistance to the attack by aggressive common chemicals
Rear wall in epoxy coated steel, designed to conform to requirements and pass the "cleanability test" according to EN12469:2000
Work surface in AISI 304L consisting of sections which are easily removable for carrying out routine cleaning and/or autoclaving sterilization procedures
Front window: electrically operated vertically sliding safety glass sash window
Re-circulating and extractor fan: SafeFAST Classic are supplied with single centrifugal fan
Servicing side glass with ports for service taps installation
Filtration: H14 HEPA/ULPA filters with an efficiency better than 99,995% MPPS (EN-1822)
Operation Condition: air cleanliness in Class ISO 3 as per ISO: EN 14644-1
The user-friendly practical keyboard and the rear-lit LCD will continuously display all required data keeping the user constantly informed of the cabinet conditions in operation
Audio-visual alarms provided for: out of range or incorrect laminar airflow velocity and frontal air barrier velocity, incorrect position of front sash-window, clogging of HEPA/ULPA filters, fan-motor malfunction, and power failure